Magnetic Resonance Microscopy. Группа авторов

Magnetic Resonance Microscopy - Группа авторов


Скачать книгу
of magnetic and electric fields within the sample.

      We start this chapter by briefly discussing the state of the art of ceramic probes in MR applications. Next, we describe several theoretical tools developed to help design MR probes with optimal performance. Finally, MRM experiments with ceramic probes are detailed and discussed.

      2.2 State of the Art

      Take-home message: Dielectric resonators have been used in microwave engineering for decades. More recently, probes for electron paramagnetic resonance (EPR) and magnetic resonance imaging (MRI) applications at several scales have been designed as well. In EPR, several theoretical tools to quantify the transmit efficiency have been developed. The engineering of high-permittivity, low-loss materials has significantly contributed to the development of these probes.

      The engineering of high-permittivity, low-loss dielectric materials has contributed significantly to the development of these probes. The high permittivity helps to confine the electromagnetic field within the resonator, while the low losses inside the material limit noise during the signal acquisition. The development of ferroelectric ceramic materials with perovskite crystalline structures has enabled the reachable permittivity to be increased while keeping the losses low [11,12].

Application B0(T) Frequency (MHz) r Imaging zone (height x diameter) Type of probe Dielectric material Exploited distribution Tuning technique
Human breast 3 128 1000 15 cm × 10 cm Volume (BaSr)TiO3 + Mg TE01δ (5 coupled disks) Interdisk gap adjustment
Wrist 7 298 80 10 cm × 10 cm Volume Water HEM11δ Capacitors (coupling loops electronic network)
Cardiac (torso) 7 298 165 Surface (BaSr)TiO3 TE01δ (array of 8 resonators)
Microscopy 14.1 600 156 24 mm × 4.8 mm Volume CaTiO3 TE01δ Overlap with copper foils
Microscopy 14.1 600 323 4 mm × 3.8 mm Volume (BaSr)TiO3 2 coupled TE01δ modes Overlap with copper strips
Microscopy 17.2 730 536 10 mm × 5.6 mm Volume (BaSr)TiO3 + Mg TE01δ Temperature
Microscopy 17.2 730 536. 10 mm × 5.6 mm twice Volume (BaSr)TiO3 + Mg 2 coupled TE01δ modes Temperature

      2.3 Modeling and Design Guidelines

      Take-home message: It is possible to model the first TE mode of cylindrical dielectric resonators, at the cost of manually writing the mode’s field expression and numerically solving a set of equations.


Скачать книгу