      Both disposable, single‐use instruments and reusable instruments are available. For veterinary practice, reusable instruments are more economical. Early instrument designs may be more difficult to adequately clean and sterilize than more current designs, a fact that should be considered when purchasing laparoscopic equipment. Instruments are available as single‐piece or as modular units, depending on the manufacturer, and can be made of high‐grade plastic or stainless steel. There are benefits for each design type. Single‐piece instruments avoid the possibility of mismatching components; however, as mentioned previously, older models may be difficult to decontaminate. Newer designs have portals that allow for cleaning of the shaft lumen. They also avoid nooks and crevices that are difficult to reach, making cleaning and decontamination easier. Modular units allow for the use of a variety of shafts and working ends with the same handles. Additionally, individual components are easier to thoroughly clean, can be sterilized separately or as a complete instrument, and can be replaced if they break rather than replacing the entire instrument.

      Regular endoscopic instruments come in 5‐ and 10 mm diameters. For most small animal applications, 5‐mm instruments are adequate. Common exceptions are the use of 10‐mm Babcock or Duval forceps for grasping thicker tissues such as the stomach wall. Mini‐laparoscopic instrumentation is now available for small dogs and cats weighing less than 10 kg. These come in 2‐, 2.3‐, and 3 mm diameter sizes, depending on the manufacturer. The available shaft length depends on the instrument diameter and should be chosen based on patient size, with shorter shafts used for small patients and longer shafts for large or obese patients. Using a shaft that is either too long or too short for the patient can be awkward for the operator and can inhibit the ability to reach tissues inside the body cavity. Short and long shaft lengths vary by manufacturer but are similar. In general, 5‐ and 10‐mm instruments range from 33 to 36 cm for the short length and from 43 to 45 cm for the long length. Mini‐laparoscopy instruments range from 20 to 30 cm in length.

      Laparoscopic forceps can be divided into three categories: grasping forceps, dissecting forceps, and biopsy forceps. Just as in open surgery, the type of forceps selected will depend on its intended use. The jaws of laparoscopic forceps can be crushing or noncrushing, straight or curved, and long or short. Some forceps are designed as either graspers or dissectors; others may be used simultaneously for both functions. In addition, graspers can be used as tissue retractors.

Photo depicts duvall and Babcock forceps (from left to right): 10-mm, double-action Duvall forceps; 10-mm, double-action Babcock forceps; and 5-mm single-action Babcock forceps.

      Source: © 2014 Photo courtesy of KARL STORZ SE & CO, KG.

      The most commonly used dissecting forceps include Kelly dissectors and Maryland dissectors (which resemble Crile hemostatic forceps) (Figure 4.5b). They are available with straight or curved jaws, but most surgeons find that curved instruments improve tissue visualization during dissection. Needle‐nosed, tapered, bullet‐nose, and micro‐dissectors are also available. For dissection around ducts and vessels, right‐angled dissectors (see Figure 4.5b) and Mixter spreader dissectors are very useful. The Mixter dissector has a longer jaw and more acute angle at the tip than does the right‐angle dissector. Dissecting forceps may be on either ratcheted or nonratcheted handles. Using ratcheted handles allows some versatility when using the same instrument for both grasping and dissecting, as the locking mechanism allows for a more stable grasp. On the other hand, nonratcheted handles allow for a smoother motion when placing and spreading the instrument jaws during dissection and avoid accidental ratcheting of the handles. When dissecting around vessels, ducts, and other fragile tissue, it is best, however, to use a nonratcheted handle.

Image described by caption.

      Source: © 2014 Photos courtesy of KARL STORZ SE & CO, KG.

Photo depicts biopsy forceps (from top to bottom): biopsy forceps without teeth, biopsy forceps with two teeth, and biopsy punch.