exchange (Rx) use. Long (300 cm) wires have become almost obsolete since the technique of balloon trapping has gained wide application to remove or insert OTW balloons or microcatheters. The only limitation of this technique is its inapplicability in very small guiding catheters (5 Fr) or when a guide extension is present (except Trapliner). Anyway, almost all wires (Abbott, Asahi, Terumo) have wire extensions that can be fixed at the distal end to increase wire length. Most wires (Abbott, Asahi, Terumo) anyway allow fitting at the distal end a dedicated extension wire. Finally, maintaining pressure with an Indeflator connected to the central lumen of the OTW microcatheter or balloon catheter allows flushing it out keeping the wire in place (Nanto technique).

Schematic illustration of the components of a rapid exchange balloon catheter.
Flexibility Flexible wires can better negotiate severe tortuosity or angulation without deformation. Shaft core material (nitinol offers greater flexibility and shape retention), core thickness (thinner core = more flexible).
Support Improved equipment delivery when hampered by angulation, tortuosity, lesion severity, calcification. Shaft core material, core thickness.
Steerability is a function of:
Torque transmission 1 : 1 transmission of torque to the tip is the ideal. Core materials are chosen for having good torque transmission, which is also improved by a thicker core with more gradual distal taper.
tip shape ability Importance increases as lesion complexity increases. Nitinol is more difficult to shape, but can be reshaped; steel is easier to shape but can be ruined by being deformed.
Lubricity Can ease wire passage in tight, calcified, severely tortuous lesions. Tip or distal segment coating with silicon, hydrophilic coating, or polymer coating: “plastic jacketed” wires are the most lubricious but also the most dangerous; when combined with a stiff tip long dissections can be inadvertently created:hydrophilic requires water for activationhydrophobic does not require water for activation, allows feedback from distal tip so that excessive friction when creating a subintimal dissection or going below stent struts can be detected.
Tendency to prolapse Can be important when negotiating angles >75° A gradually tapering core with a smooth transition toward the tip improves support and tracking around bends; abrupt tapers and floppy cores are more likely to prolapse.
Visibility The level of visibility becomes more important in obese patients or when angled working views are required. Lubricious polymer‐coated nitinol wires can be difficult to see, platinum, steel, or tungsten markers at distal tip.
Tactile feedback Provides the operator with essential nonvisual information, allows “palpation” of the lesion at the distal wire tip. Hydrophylic wires provide poor tactile feedback; hydrophobic wires provide more feedback.
Tip stiffness A soft gentle tip is essential for all “workhorse” wires; to reduce the risk of vessel trauma; stiffer tips are required for dedicated CTO wires. More gradual distal taper, distal core material (e.g. high tensile steel).

Скачать книгу
Librs.Net
Product name Core material Rail support Radiopaque tip length (cm) Tip type Tip style Tip tapering Tip stiffness Tip load (g) Tip coating type
Workhorse wires
Hi‐Torque BMW Universal Elastine Nitinol Moderate 3 Polymer/ Spring coil Shaping ribbon Non‐tapered (0.014″) Soft 0.7 Hydrophilic
ChoICE Floppy Stainless steel Light 3 Polymer/ Spring coil Core‐to‐tip Non‐tapered (0.014″) Soft 0.8 Hybrid (distal 3 cm uncoated)
Runthrough NT Stainless steel/ Nitinol Light 3 Stainless steel coils Shaping ribbon Non‐tapered (0.014″) Soft