Interventional Cardiology. Группа авторов

Interventional Cardiology - Группа авторов


Скачать книгу
in typical catheterization laboratory patients without a specific protocol for sedation.

      FFR and iFR‐Pullback is used to determine the resting pressure loss caused by each individual stenosis and can be particularly useful when multiple coronary stenoses are present or if there is additional diffuse disease that may modulate the likelihood of hemodynamic improvement after coronary intervention. Performing a pullback assessment provides useful information that can guide an interventional strategy [43]. However, there are significant differences in how this information can be used.

      This issue can be resolved by measuring coronary wedge pressure, but this is not a practical solution for typical clinical practice. Instead, FFR pullback should be performed to measure pressure gradients throughout a vessel, and it is recommended that the lesion with the greatest pressure‐gradient is treated first. After this, it is recommended that the FFR measurement and Pullback is repeated again, as there may be a change in significance of the residual stenoses. This is pertinent, as a distal stenosis not previously hemodynamically significant may become significant once a proximal stenosis is relieved.

Schematic illustration of IFR Pullback can be used to predict post-PCI iFR values.

      When performing an iFR‐Pullback, continuous beat‐to‐beat iFR values are calculated and plotted as continuous line. The shape of the line describes focal and diffuse disease. As the pressure sensor crosses a stenoses, a step in pressure trace will be observed.

      Virtually removing a given stenosis can be performed by studying the pressure pullback trace and the numerical values measured. A given stenosis can be discounted from the trace, and the residual pressure gradients will remain to compute a post‐PCI iFR value [22,23]. As the computer algorithms have iterated, it has become easier to remove different segments from the pressure trace to model the hemodynamic effect of a given intervention. In each case, the stented segment would remove the pressure loss in the area of interest, but the residual disease continues to cause the pressure loss. It should be noted that stent expansion may be not be perfect and if a post‐PCI result does not match the predicted iFR result, then attention should be paid to stent optimization. Repeat iFR pullback will reveal relevant residual pressure gradients.

      The speed of wire movement will determine if the step‐change is very shallow or very steep. When a consistent pullback speed is utilized, steep steps tend to represent focal disease while more shallow changes represent diffuse disease. Even in focal disease, the pressure steps can be small or large; a single stenosis may in fact cause several different areas of pressure change. If the wire is moved too quickly through the lesions such granularity may be lost, but the important aspect of observing the step will remain true.

      The iFR‐Pullback trace is made up of two separate lines: (i) the dotted line, which represents the raw iFR value at the given location, and (ii) the solid line which represents a processed value that prevents the iFR‐Pullback line from falling below what has already been plotted. In many cases, both lines overlap and may not be visible. In some cases, it is possible to see that the dotted line falls significantly at the outlet of a stenosis before returning to the solid line. This represents a pressure‐recovery phenomenon as the sensor moves through a stenosis. It should also be noted that if there is a whipping artefact or ectopic, then the solid line may move artificially. Since it will always continue from the highest value, then these events can make the pressure trace less reliable for calculation of post‐PCI iFR results. In these situations, the dotted line should be assessed as this will return to the true live value.

       Allow a steady trace before pullback: When performing an iFR pullback assessment, a short period of time should be allowed from the moment the pullback is activated to when the pullback is started. This will enable a steady line for pullback to be displayed. If there is movement of the line before pullback has started, it is advisable to cancel the pullback and start again.

       Pullback the wire without opening the pressure port: the pressure wire can be withdrawn without altering the O‐ring or opening a port.

       Continuous fluoroscopy should be used when using co‐registration systems: If this is not available, a brief fluoroscopic image stores should be made when the pressure steps are noted on the pullback trace. This will aid the interpretation of small or steep changes in pressure and help the operator co‐localize the pressure changes with angiographic appearances.

       Always return to the ostium: The pressure sensor should be moved to the ostium of the vessel as part of the pullback. This means that any drift can be identified. The pullback systems automatically reformat the pressure trace for small degrees of pressure‐wire drift which assists in interpreting the data collected.


Скачать книгу