A Course in Luminescence Measurements and Analyses for Radiation Dosimetry. Stephen W. S. McKeever

A Course in Luminescence Measurements and Analyses for Radiation Dosimetry - Stephen W. S. McKeever


Скачать книгу
(e.g. MgB4O7:Mn, MgB4O7:Dy, and MgB4O7:Cu), Al2O3 (e.g. Al2O3:C; Al2O3:Si,Ti, and Al2O3:Mg,Y) and CaSO4 (e.g. CaSO4:Tm and CaSO4:Dy). Other, less well-known and utilized materials have often been described in the published literature, including natural minerals (e.g. calcite, fluorite, quartz). OSLD is dominated by Al2O3:C and BeO, while RPLD is dominated by Ag-doped alkali phosphate glass, and Al2O3:C,Mg.

      The material properties that are important in personal dosimetry are a high sensitivity (large signal for a given dose), a linear dose-response relationship, angle independence, and energy independence. The latter is difficult to achieve at low radiation photon energies. However, the important property in this regard is that the material responds to low-energy photons in the same way as does human tissue – a property known as tissue-equivalence. The most important characteristic for tissue equivalence is the effective atomic number Zeff of the material. Here, materials such as LiF (Zeff = 8.14), BeO (Zeff = 7.4) and Li2B4O7 (Zeff = 7.13) have advantages over other materials since these values are close to that of tissue (Zeff = 7.4). However, tissue equivalence is not essential as long as the response at low energies is well defined such that suitable corrections can be made (for example Al2O3:C with Zeff = 10.2).

      1.3.2 Medical Dosimetry

      Modern advances in radiation medicine – in radiodiagnosis, radiotherapy, and interventional radiography – are now presenting dosimetry challenges that did not exist previously. For example, the movement toward the use of charged particles (protons and carbon ions) in radiotherapy has resulted in new requirements for luminescence dosimeters compared to the dosimetry of high-energy photons. Known responses to charged particles means careful calibration of the dosimeters in charged-particle fields. This consideration becomes especially significant beyond the Bragg peak for energy deposition by charged particles. Here, the linear energy transfer (LET) of the particles increases rapidly as the particles lose energy and slow down. Knowledge of the behavior of dosimeters in such regions, where the physical dose is rapidly decreasing but the relative efficiency of biological damage (compared to high-energy photons) is rapidly increasing, is a research area of growing importance. Secondary neutron production is also a feature of charged-particle irradiation and the response of the dosimeter in neutron fields also needs to be established.

      Sophisticated intensity-modulation techniques in radiotherapy, whether using photons or particles, also create new challenges for dosimetry. This is due to the production of sharp dose gradients where the therapists try to protect healthy tissue surrounding the tumor as much as possible. It is also necessary to understand the response of the dosimeter in either steady or rapidly pulsed radiation fields.

      The luminescence dosimeters may be used in vivo or ex vivo. Their small size enables their innovative use internally as the patient is treated (for example, in brachytherapy). The dosimeters may also be used externally to measure entrance and exit doses on the surface of the patient without disturbance of the radiation field.

      Luminescence (particularly OSL) has also been used as an imaging technique in radiodiagnosis (where it has long been known by its alternate name of photostimulated luminescence, PSL). The sensitivity and speed of readout of the stimulated luminescence signal has given radiologists the ability to reduce radiation doses to patients and yet provide high-resolution images to aid diagnosis. However, the use of OSL in this way is not dosimetry. The actual dose to the patient is still determined by other methods.

      In interventional radiography the real-time dose to patients (and surgeons) during surgical procedures can be significant. Adaptation of OSL techniques using fiber optics can provide dose measurements in real time, without disturbing the x-ray image.

      In all of the above applications, the preferred properties of the needed dosimeters include tissue equivalence, high sensitivity, small size, rapid response, and known responses as functions of dose, angle, photon energy, particle LET, and neutrons.

      1.3.3 Space Dosimetry

      Dosimetry (for astronauts and equipment) in space environments is one of the most challenging applications for luminescence dosimetry. The space environment consists of an extremely wide range of highly energetic charged particles originating from cosmic rays (CR), solar particle events (SPE), and the radiation belts trapped by the Earth’s magnetic field (Earth’s radiation belts, ERB). Additionally, once these particles interact with the spacecraft structures, the astronauts’ space suits, and the astronauts’ bodies, secondary charged particles and neutrons are also produced. Personal and environmental dosimeters for use in space therefore need to respond to an extremely wide-ranging set of particles, each with its own efficiency of luminescence production. The dosimeter’s response to each particle type needs to be well-known and calibrated.

      1.3.4 Retrospective Dosimetry

      In one sense, all passive methods of radiation dosimetry are retrospective in that they produce a measurement of dose only after the dose has been delivered. That is, they integrate the total dose received by the dosimeter since the dosimeter’s last reading. (This is opposed to active dosimeters, which record the dose or dose rate in real time during the exposure period.) However, the term “retrospective” is here reserved for assessing the dose received by an individual who may have undergone an acute exposure during a radiation accident and who may not have been wearing a conventional dosimeter at the time. During radiation accidents (i.e., accidental exposure to a radiation source or contamination by radioactive pollutants), members of the public may be exposed, and since it is not


Скачать книгу