Encyclopedia of Glass Science, Technology, History, and Culture. Группа авторов

Encyclopedia of Glass Science, Technology, History, and Culture - Группа авторов


Скачать книгу
resolution is improved in this way, the disadvantages include an enhanced susceptibility toward contamination of the scan field as well as a practical inability to perform X‐ray spectrometry because the electron energy will be insufficient to induce X‐ray emission from the sample.

      2.2 Chemical Analyses

      Characteristic X‐rays excited by high‐energy electrons can be recorded either by energy‐dispersive X‐ray spectroscopy (EDXS) or by wavelength‐dispersive X‐ray spectroscopy (WDXS). Stemming from electronic recombination, the X‐rays generated fingerprint the electronic structure of the excited element, thus making quantitative analyses possible. Soft X‐rays (e.g. emitted by light elements) are subject to absorption on their way out of the sample, however, so with decreasing atomic number, the emission of Auger electrons counteracts that of X‐rays. Hence, quantification for elements with atomic numbers smaller than 13 requires extraordinary diligence.

      In EDXS, a semiconductor detector is used to count electron‐hole pairs generated by the impinging X‐rays. The spectral resolution is typically on the order of 130 eV, so in a few cases overlapping emission lines cannot be clearly separated: this will, for instance, apply to the Ti Kα1 and Ba Lα1 at about 4.51 and 4.47 keV, respectively. This problem is less serious with WDXS (with a spectral resolution of ≈10 eV), where Bragg’s law of selective reflection forms the basis of detection in a proportional counter. Whereas EDXS is a parallel‐registering technique, WDXS (which is very similar to electron microprobe analysis [EMPA]) is a serial technique, making it more time consuming and less commonly used than EDXS.

      2.3 Application to Glass Ceramics

      To illustrate the information depth one can gain with that technique, the aforementioned MAS glass ceramics will be selected. In this sample, precipitation of zirconia and spinel is known from X‐ray diffraction (XRD) experiments, but this technique is inappropriate for investigating the micro‐ and nanostructure because it gives information about the sample as a whole, with a low spatial resolution defined by the X‐ray spot size. Techniques such as SEM or TEM must be used instead, and they can in addition yield the composition of the residual glass and the structure of the crystals – in this case whether zirconia is the tetragonal, cubic, or monoclinic polymorph.

Photos depict the scanning electron micrographs of the MAS glass-ceramic sample: resolution differences between 10 kV (a) and 30 kV (b) acceleration voltages and BSE micrograph taken at 10 kV at higher magnification (c) showing the bright (1) and dark (2) areas for which the EDX spectra of (d) were acquired.

      To check this conclusion, EDXS analyses were made in the two areas indicated in Figure 2c. The spectra are presented in Figure 2d. In both, a small carbon peak is an artifact of the sample preparation. Besides the most intense oxygen and silicon peaks, the other peaks are readily assigned to the other elements, but a clear separation of the Y‐Lβ and Zr‐Lα peaks is not possible owing to the limited energy resolution of the EDX detector system. The better separated Y‐K (at ≈14.9 keV) and Zr‐K lines (at ≈15.7 keV) would be more suitable, but a higher acceleration voltage would be needed to access that higher‐energy part of the EDX spectrum with its ensuing loss of image resolution.

      Nevertheless, these EDX spectra suggest that the brighter areas likely are silica rich, most probably depicting the residual Y‐bearing glass. Conversely, the darker areas should consist of spinel, which is Si‐ and Y‐free, as confirmed by analyses made in the “darker” area of the sample where O, Al, and Mg are found with a higher Al/Si ratio than in the other spectrum. The reason why certain amounts of Si and also of Zr (but not of Y) are nonetheless detected is that a certain interaction volume is also excited below the surface within the sample (cf. Figure 1). As a result, the lateral resolution is worsened, and the probed area does not include only the “dark” region 1 that is visible in Figure 2c, yet also parts of the sample bulk below. In conclusion, like any technique, SEM and EDXS have their own limitations. Other methods having a higher resolving power must be used to overcome them.

      3.1 Conventional Observations

      Analytical (scanning) transmission electron microscopy [(S)TEM], currently is the best method for both imaging and determining chemical compositions [18] whenever the relatively large excitation volume of SEM becomes problematic. The price to be paid is the necessity to prepare samples as thin as a few 10 nm, without preparation artifacts, to make them transparent to electrons. The task is difficult, but over the years a wide range of techniques have been developed for this purpose, including mechanical grinding, ion beam etching, preparation of replica, ultramicrotomy, and FIB machining of lift‐out TEM lamellae [19]. In this respect one should distinguish between samples taken from the bulk and cross‐sectional samples cut out from specifically targeted areas (e.g. if a specific source of devitrification in a glass has to be assessed for further analysis) for which FIB machining is particularly well suited.

      In its principle, a transmission electron microscope is amazingly similar to a light microscope: the electron source corresponds to the lamp; beams are formed by electromagnetic lenses (which are true zoom


Скачать книгу