Emergency Medical Services. Группа авторов
Jarvik 2000 (Jarvik Heart)
Third generation (centrifugal flow)
Berlin Heart INCOR (Berlin Heart AG)
CentriMag (Abbott Laboratories)
HeartMate III (Abbott Laboratories)
HeartWare HVAD Pump (Medtronic)
Fourth generation
HeartAssist 5 (ReliantHeart Inc.)
Axial pumps use a corkscrew and the Archimedes principle against gravity. The inflow and outflow pumps are in line with the impeller, resulting in a smaller size pump. In contrast, centrifugal pumps have the inflow and outflow cannulas at right angles to the flow. Right angles allow for less suction, which can decrease the risk of the ventricle collapsing around the inflow cannula or distortion of the interventricular septum. Both can result in right ventricular failure [11]. These continuous‐flow devices are found in 90% of patients with LVADs. Continuous‐flow devices have been shown to produce superior organ perfusion than first‐generation, pulsatile, devices [13].
Third‐generation LVADs represent a further technological step forward. They can use continuous flow, axial flow, or centrifugal pumps. The impeller is driven by electromagnets. This results in no contact between the impeller and the sides of the pump. Benefits include less trauma to blood components and less thrombus formation. The devices are also quieter and can last longer [11]. Fourth‐generation LVADs, currently in testing and trials, are exploring further advances in technology, including wireless monitoring and elimination of the driveline. This would remove the cabling from the pump, which must travel through the skin in order to connect to the power source. As driveline infections are a major source of LVAD complications, driveline removal could result in significantly less morbidity [11].
Figure 11.2 HeartMate II LVAD. LVAS, left ventricular assist system.
Source: Reproduced with permission of the Thoratec Corporation.
The results of these advancements in technology are devices that allow the patient to leave the hospital and function at home. Prior to discharge the patient and family are given extensive training on the operation and maintenance of the device and how to troubleshoot problems and alarms. The patient is followed by a hospital team and is given written instructions for EMS practitioners, which outline the device operation, emergency interventions, and hospital contact information [14].
LVAD complications can be divided into two categories: device problems and patient problems (Box 11.2). The most common problems consist of neurologic events, bleeding, and cardiac arrhythmias. Neurologic events include acute strokes and transient ischemic attacks. Thrombotic and hemorrhagic events can occur. The incidence of stroke in VAD patients has been reported ranging from 8% to 25%. The risk is increased for patients with stroke histories and those who have had device‐related infections [11].
The most commonly experienced forms of bleeding include epistaxis, gastrointestinal bleeding, and hematoma formation. Bleeding can result from trauma to blood components, from acquired von Willebrand disease, or from iatrogenic anticoagulation [15]. Most patients are given anticoagulants or antiplatelet drugs to reduce the risk of thrombus formation [16]. LVAD patients are also at increased risk of arrhythmias. Patients may have atrial fibrillation, often as a result of underlying disease. The LVAD will provide left ventricle support, but the loss of atrial “kick” may affect right ventricular function. LVAD patients may also suffer from ventricular arrhythmias. These arrhythmias may result from underlying disease, from irritation of the myocardium by the device, or from ventricular collapse or septal deviation from excessive pump function. Some patients may require an implanted cardioverter defibrillator [11, 16].
Infection is the most common complication, with infection rates ranging from 18% to 59% among LVAD patients. Infection is second only to heart failure as a cause of mortality in these patients. Infections can present at the surgical site, the driveline, the pump pocket, or the pump itself in the form of endocarditis [17].
Box 11.2 Complications encountered in LVAD patients
LVAD‐specific complications
Suction event
Pump thrombosis
Pump complications: failure, stoppage, driveline damage
LVAD‐associated complications
InfectionDevice‐related (i.e., endocarditis)Device‐specific (i.e., driveline and VAD pump pocket infection)Non‐LVAD infection (i.e., urinary tract infection, pneumonia)
Bleeding (i.e., gastrointestinal bleeding)
Cerebrovascular pathology: ischemic or hemorrhagic stroke
Hemolysis
New right ventricular failure
Dysrhythmia
Aortic regurgitation
Device‐specific problems can manifest as device failure (fortunately rare) or from battery or cable connection issues. Suction events can occur when there is not enough volume in the left ventricle to support the speed of the pump. This causes the intake cannula to collapse and subsequent ventricular arrhythmias [11]. LVADs in place for a long time can become dislodged, resulting in incomplete left ventricle emptying, right ventricular failure, and arrhythmias. LVADs may also have thrombotic complications, causing problems ranging from dyspnea to cardiogenic shock [18, 19].
While EMS clinician interactions with LVAD patients may be infrequent, these patients are high acuity and tend to have high rates of hospital admission [20]. It is beneficial for EMS systems to be aware of LVAD patients in their service areas, and have device reference cards and VAD specialist contact information accessible [21]. Hospital policies may indicate that a VAD specialist be sent to the scene to evaluate the device in the event of a problem. This situation could then result in a delay in patient transport while the VAD specialist is en route. If the patient is having a time‐sensitive medical issue not related to the device, a medical oversight decision may need to be made regarding transport. For example, perhaps a portable LVAD patient is having an acute stroke or GI bleed and no LVAD issues, but the EMS crew is not familiar with the LVAD. The medical oversight physician will need to weigh the risks of delay of transport while waiting on the VAD specialist to arrive versus the risk of an LVAD complication during the EMS transport.
The LVAD patient in distress might be having an issue with the device, an exacerbation of the underlying cardiac disease, or an unrelated medical event. The initial EMS assessment should be to determine if the issue is LVAD‐related or not (Figure 11.3). If the event does not seem to be LVAD‐related, then local protocols or medical oversight should be consulted for further guidance. The next step is to determine the type of LVAD involved. The patient and caregiver should have device information available. This information should include whether the patient can receive electrical therapy and whether or not CPR can be performed. Obviously, these questions need immediate answers [11, 21].
EMS personnel must determine if the device provides pulsatile flow or continuous flow. A patient with a pulsatile flow device should have a palpable pulse and blood pressure. Pulsatile‐pump LVAD failure requires