Antisepsis, Disinfection, and Sterilization. Gerald E. McDonnell

Antisepsis, Disinfection, and Sterilization - Gerald E. McDonnell


Скачать книгу
and cleaning indicators that use artificial test soils inoculated onto a surface to test (generally by visual inspection) physical removal during a cleaning process or cycle. Mechanical indicators play an important role in the parametric release of a product or process as an alternative to the use of chemical and biological indicators for routine monitoring of sterilization processes (see section 1.4.2.4).

Class Type Description
1 Process indicators Indicate exposure to minimal process conditions; used to differentiate exposed from unexposed items (e.g., autoclave tape)
2 Indicators for use in specific tests Indicate that a specific process is obtained, which is linked to the sterilization process (e.g., a Bowie-Dick test indicates the adequate removal of air from a prevacuum steam sterilizer)
3 Single-parameter indicators Indicate a change on exposure to one parameter (e.g., concentration of a biocide or temperature).
4 Multiparameter indicators Indicate a change on exposure to at least two parameters
5 Integrating indicators Indicate a change on exposure to all the critical parameters of a given process (e.g., ethylene oxide sterilization with temperature, biocide concentration, relative humidity, and time)
image

      The concept of parametric control (or release) as a method to verify the effectiveness of a biocidal process is based on the understanding of all the key physical parameters that can affect its success or failure. Although theoretically this could be applied to any disinfection or sterilization process, it is generally restricted to well-characterized sterilization methods, including steam, dry heat, ethylene oxide, and ionizing radiation. An example is steam sterilization. The efficacy of steam is affected by the temperature and time, but also by the presence of air (see section 5.2). These parameters are reasonably well understood and can be physically measured (by using mechanical indicators [see section 1.4.2.3]) to ensure that the correct conditions have been met during a given steam sterilization cycle. In addition to monitoring these conditions, a series of in-process tests and controls are also conducted to provide further assurance that the sterilization process has been efficient. However, the concern with parametric release as an alternative to biological monitoring is in the control of other variables that can affect the effectiveness of the process. In the case of steam, these include the quality of the steam (see section 5.2), variations in the load being sterilized, and in the case of reusable devices, if the cleaning process has been sufficient prior to sterilization (see section 1.4.8). In most cases, disinfection and sterilization processes are routinely tested and monitored using a combination of biological and chemical indicators in parallel with mechanical indicators for parametric control.

Title Summary
ISO 15882 Chemical Indicators—Guidance on the Selection, Use, and Interpretation of Results Guidance for the selection, use, and interpretation of results of chemical indicators used in process definition, validation, and routine monitoring and control of sterilization processes
ISO 11140-1 Sterilization of Health Care Products—Chemical Indicators—Part 1: General Requirements General requirements for production, labeling, test methods, and performance characteristics of chemical indicators to be used in the validation and routine monitoring of sterilization processes
ISO 11140-3 Sterilization of Health Care Products—Chemical Indicators—Part 3: Class 2 Indicators for Steam Penetration Test Sheets Specific requirements for class 2 steam penetration test indicators
EN 867-1 Non-Biological Systems for Use in Sterilizers—Part 1: General Requirements General requirements for indicators that are used to monitor the presence or attainment of one or more sterilization process variables
ANSI/AAMI ST60 Sterilization of Health Care Products—Chemical Indicators—Part 1: General Requirements Requirements for chemical indicators intended for use with sterilization processes employing steam, ethylene oxide, irradiation, or dry heat

      Other specific methods are used to evaluate the efficacies of biocidal processes and products. These are generally used due to restrictions on the cultivation of various microorganisms under laboratory conditions. Microscopy or other detection (biochemical and genetic) methods for the presence of microbial contaminants have been used; however, these methods can identify the presence of an organism but may not necessarily detect its actual viability. A specific example of the use of microscopy is in the determination of the viability of protozoan (oo)cysts or helminth eggs. In the case of protozoa, the viability of cysts can be determined in vitro by suspension in a specific medium under controlled conditions (including temperature) to cause the cysts to excyst and release their vegetative forms (see section 1.3.3.4); this can be monitored microscopically and is considered a reasonable indication of cyst viability. Cell culture methods (using mammalian cells) have also been developed to determine cyst viability and are also widely used to culture viruses and some bacteria. In many of these cases, in vitro methods are not sufficiently developed to determine microbial viability, and the use of in vivo (animal) models is required. An example of this is the case of prions (see section 1.3.6), which are proposed to be infectious proteins; although the biochemical