Spectroscopy for Materials Characterization. Группа авторов

Spectroscopy for Materials Characterization - Группа авторов


Скачать книгу
measurements can be easily achieved by a setup such as in Figure 3.5. To achieve this aim, one repeats each measurement in the two pump polarizations, parallel (para) or perpendicular (ortho) to the gate, respectively, and calculates the fluorescence anisotropy r(t):

      (3.18)equation

      The actual design of a FLUC setup involves a relatively complex optimization problem involving multiple geometrical parameters to find the best compromise between signal intensity, noise rejection, and time resolution. While this topic cannot be fully covered here, a significant example is given by the geometry of the telescope directing the fluorescence into the nonlinear crystal. In fact, the focal lengths of the two parabolic mirrors used in Figure 3.5 must be chosen in such a way that the fluorescence enters the BBO within the acceptance angle for the SFG process. If n o is the ordinary refractive index, and θ m is the phase matching angle, the SFG acceptance angle is given by [47]:

      (3.19)equation

      The need to fulfill this condition imposes relatively tight limitations on the possible geometrical parameters of a setup such as in Figure 3.5. Similar considerations can be done for many other relevant geometrical parameters, the optimization of which constitutes an important aspect in the design of any FLUC setup [47].

      3.4.3 FLUC: Data Analysis and Interpretation

      Data collected in a FLUC experiments are usually single‐wavelength traces. The treatment and correction of the data involve, as for the TA, the definition of a common zero time for all the kinetics, the GVD correction, and an assessment of the temporal resolution. A useful technique to measure the temporal resolution is the measurement of the temporal evolution of a Raman signal of a pure solvent. Because Raman scattering is essentially instantaneous, the temporal width of the Raman signal corresponds to the IRF of the experiment and it can be used to calibrate the experiment. After these preliminary procedures, the results can be treated by GA methods similar to what discussed above for TA, although FLUC data are generally much easier to interpret, inasmuch as they are much less affected by spectral congestion issues, and only contain one type of signal, that is spontaneous emission. In fact, many FLUC signals essentially consist in a single time‐dependent fluorescence band, or a few with no mutual overlap. In this case, FLUC data are often simpler to analyze by direct band integration methods (e.g. moment analysis), leading to very straightforward interpretations. For example, one popular application of FLUC is probing solvent relaxations around a chromophore through the observation of picosecond and femtosecond dynamical Stokes shift induced on its fluorescence band. This approach has proven very valuable to analyze solvation dynamics in water [3], and hydration dynamics around proteins, a problem of great biological importance [4, 44].

      3.4.4 Kerr‐Based Femtosecond Fluorescence Spectroscopy

      The temporal resolution is determined by whatever is shorter between the cross‐correlation between excitation and gate and the cross‐correlation between the excitation and the intrinsic response function of the Kerr medium. In fact, if the Kerr medium reacts slowly to the gate modifications, then the temporal resolution is dominated by the latter. For example, with an excitation of ≈50 fs at 475 nm and a gate of 40 fs, the time resolution is ≈120 fs using a 1 mm‐thick fused silica plate as a Kerr medium [49].

      Femtosecond time‐resolved methods are not limited to electronic spectroscopies, but can also be extended to meet the advantages of vibrationally sensitive techniques, such as infrared absorption and Raman. In this regard, this section addresses ultrafast stimulated Raman spectroscopy, which is one of the most powerful methods to combine the structural sensitivity and chemical selectivity of vibrational methods with the power of femtosecond time resolution [50–55].

      3.5.1 The Experimental Method

      Classic Raman spectroscopy exploits the Raman effect, discussed in detail in Chapter 5 of this book, that is the inelastic scattering of light associated to the generation or annihilation of vibrational quanta of energy ω V during the interaction of the light beam with the system under study. A typical steady state Raman experiment interrogates the sample by a laser beam at ω, and analyzes the spectral distribution of photons scattered in all directions. In the most common case (Stokes Raman scattering), part of the scattered radiation is found to be red‐shifted to ω′ = ωω V, because an energy ω V has been spent to excite phonons of this frequency within the sample. Thus, spectral analysis of the scattered radiation reveals a series of peaks at various ω V reporting on the vibrational mode structure of the system. Vibrational frequencies being much smaller than visible ones, the exciting beam must be narrow enough in frequency to avoid spectral overlap between ω and ω′. Therefore, the vibrational spectral resolution is of the order of the FWHM of the Raman pump beam. Raman spectroscopy is a powerful probe of the vibrational mode structure of any physical system and displays a sensitive dependence on both electronic and molecular structure. Thus, it can be used for accurate chemical recognition and to reveal, with high specificity, any structural changes of the system at study.

      On these grounds, femtosecond time‐resolved Raman (TRR) experiments can be conceived as a powerful probe of excited state dynamics of any photoexcited physical system, overcoming the lack of structural sensitivity typical of absorption and fluorescence spectroscopies. In particular, Raman spectra being characteristic of any given state of the system, one should expect strong changes of the vibrational mode pattern whenever the system transitions to an excited state, and subsequently evolves. Following these changes by TRR can be used to extract detailed information on any structural changes or photochemical processes initiated by photoexcitation.


Скачать книгу